| Country of Origin | Made in India |
| Size | A6 |
| Usage | Stem Sterilization |
| Colour change | Yellow to Black |
| Disposable | Yes |
| Operationg temp | 15 MINUTES AT 121 DegreeC AND 7 MIN FOR 134 DegreeC |
| Single use | Yes |
Pre assembled for single use test, packs designed to evaluate the performance of the air removal system of pre-vacuum sterilization operating at 134° c for 3.5 minutes & 121° c for 15 minutes.
COA OF BOWIE DICK TEST PACK
INDICATIONS FOR USE
THE AIR REMOVAL BOWIE-DICK TEST PACK IS FOR DAILY MONITORING OF
PRE-VACUUM STEAM STERILIZERS OPERATING AT
121°C FOR 15 MINUTES 134°C FOR 7 MIN. BOWIE DICK TEST PACK FOR
STEAM STERILIZATION PROCESS (120 MM X 120 MM X 16 MM) ACCORDING
TO ISO 11140-1 STANDARD.
COLOUR CHANGE
THE INDICATOR WILL CHANGE TO BROWN/BLACK WHEN EXPOSED TO
STEAM STERILIZATION CYCLE WITH ACCEPTABLE VACUUM CYCLES.
DIRECTIONS FOR USE
THE BOWIE-DICK TEST IS CONDUCTED EACH DAY IN AN EMPTY STERILIZER
PRIOR TO PROCESSING ANY LOADS.
1.RUN A SHORTENED CYCLE (OMITTING THE DRYING PHASE) TO
PROPERLY HEAT THE STERILIZER.
2. PLACE THE TEST PACK ON THE BOTTOM RACK OF THE STERILIZER
NEAREST TO THE DRAIN.
3. RUN THE STERILIZER FOR 15 MINUTES AT 121°C AND 7 MIN FOR 134°C.
4. REMOVE THE TEST PACK FROM THE STERILIZER AND ALLOW THE PACK
TO COOL. DO NOT PLACE ON A COLD SURFACE. PULL EASY OPEN ZIP TAB
ACROSS BOX FRONT TOO OPEN.
5. REMOVE INDICATOR SHEET FROM THE PACK AND DISPOSE REMAINING
PACK MATERIALS.
Additional Information:
- Item Code: SIG3V
- Delivery Time: 1-2 Days
- Packaging Details: Poly Packing